• No labels

1 Comment

  1. RF03
    Tech: Richard Freeman
    Date: April 13, 2022

    Takeoff: 1735
    Landing: 2258

    Pre-flight:
    2DS - Relative Humidity at 18% all element voltages above 1.5V
        - Applied mask to vertical channel
     

    Flight:
    Could not start cameras after takeoff. Desktop button consistently showed an application error splash screen. Attempted to run via terminal twice without success. terminal output loop an error message with an error code -13. (Catherine and Chris were able to fix later in flight.

    1808 - 2DS generating many false particles on the verticle channel. Applied mask to stuck bit in vertical array.


    1824 - DPL and DPR went out of balance. cycled power without result.
    1840 - Reset USHAS-G software
    1844 - Reset power to down camera at L2 rack PoE box.
    1931 - Service mirror light indicator came on for DPR
    2018 - Cycled power on both dew pointers. Lost all DSM variables for one minute as a result. (known grounding issue related to PSF?) 
    2020 - Reset UHSASG software
    2105 - Both Dew Pointers service mirror light turned on. Per Cory, decided to leave both off rather than risk a reset of other flight variables.