Mask application is being either ineffective or chaotic.
Relative Humidity (RH) = 6% to 18% (fluctuating the whole flight)
Between 2310 and 2322 the "KBytes Received" value changed to a negative number. It is growing more positive.
Between 0023 and 0037 flew through a very wet cloud which cleared debris from the optics. Vertical Element Voltages went from 0.3(85) and 0.9(127) to well above acceptable ranges.
By the end of the flight - H Particles =1678800 V Particles =18181893
I took a handful of photos of these events. I'll make them immediately available to Josh via email and put them on the Tech drive in the 2D-S folder. Let me know if you want me to share them with you via email.
1 Comment
David Allbee
Tech - Dave Allbee
4/15/2022
Taxi - 1821
Takeoff - 1827
Landed - 0243
No SATCOM issues
CDP appears to be functioning well
DPL, DPR, VCSEL not tracking one another.
DP sensors likely need to be cleaned.
0056 - DPR Service Mirror came on.
2D-S
Mask application is being either ineffective or chaotic.
Relative Humidity (RH) = 6% to 18% (fluctuating the whole flight)
Between 2310 and 2322 the "KBytes Received" value changed to a negative number. It is growing more positive.
Between 0023 and 0037 flew through a very wet cloud which cleared debris from the optics. Vertical Element Voltages went from 0.3(85) and 0.9(127) to well above acceptable ranges.
By the end of the flight - H Particles =1678800 V Particles =18181893
I took a handful of photos of these events. I'll make them immediately available to Josh via email and put them on the Tech drive in the 2D-S folder. Let me know if you want me to share them with you via email.