...
Issue | Action Taken | Future Actions |
---|---|---|
USB sticks died | Switched to Pocketec, Compact Flash, New stick | Replace all sticks with new ones recently purchased |
GPS timing odd | Widened range of acceptable time lag | Upgrade software on all receivers (3.0-3.7 -> 3.9) |
Qsoil reading 0x8000 | none | Fix - connector?Check connector at tear-down |
CSAT3 broken upon installation | Repaired at CSI – bad coax cable in head/box | Unknown - new handling procedure? |
XBee mote not reliable over 3m distance | Changed to serial mote (tried changing power) | Change to all Bluetooth motes? |
TRH went to large values | eio power cycle | Unknown |
TRH fan controller died | Replaced | Check; see if systematic issue for all |
Lost data connection when eol-rt-data changed | eol-rt-data configuration restored crontab script fixed | SIG checks with Gordon (he tells them when using?) SIG avoids "snapshots" |
Odd Gsoil at ehs | none | Don't install in darkhorse shadow! |
EC150 bad data for a week | Cleaned bird scat | Install bird diverters? |
tsoil.bao mote data intermittent | none; came back | Unknown |
Mast issues: guy wires; winch; clamp positions; tight nubs | Various in-field adjustments | Systematically check each component for all masts |
AC powering failed (probably loose connections) | Changed to solar power | Test this option (needed in future) |
Picklefork Tsoil died (then revived!) | none (had installed duplicate picklefork) | New picklefork design |
...