- trh.0,2m.rel: data are bonkers, with T=-61, RH=0. Turned out that the SHT was loose inside the Gortex shield. Reseated SHT and started working.
- trh.10m.init: exactly the same issue. Will climb tomorrow to fix.
- csat.10m.uconv: was plugged into the wrong port (usb4 rather than usb6). Fixed, but data only coming in at 1sps. Need to reconfigure sampling rate.
- csat/irga.1.5m.lconv: was not getting power. Swapping cables with the csat.3m blew the power front panel's 1A fuse. Changed power bank and was able to bring everything up, though the EC150 LED never went green. Next step is to replace the EC150 head. I think we have a spare.
- trh.15m.lconv and trh.20m.lconv. Should be getting power and jumpers set to RS232, but no data or response to commands. I suspect that the SHTs were not installed in the housings. We'll check this tomorrow as well.
I did find some cockpit plot range issues, that I fixed, and others that I didn't (don't know why not).
...