- ue3m - sonic + gas lights are red. getting messages from sonic, but uvwtc, co2/h20, tirga/pirga are all nans. → after power cycle w/ pio lights are green + reporting winds + gas
- ue10m - same behavior as ue3m, fixed same way
- c5m - lights are green but no data (at all) on port. power cycle didn't help. note on ec100 that the pelican case it came out of had been labeled 'needs output reformat' → set output format using ECmon, fixed
- c10m - red for sonic, green for gas and power. gas data ok, nans for winds (but then why is gas data also missing on qctables?). sonic diagnostic is 16, or 'acquiring' → took sonic out of box to see if something was in the way, but no improvement. William and I tried a different sonic head, but it still showed up as red and didn't give winds. Considering this, and considering that in theory all the sonic heads were checked during cals and it's the boxes that are untested, we think the next thing to try would be swapping the EC100 box, which we'll try tomorrow.
- c15m - see above → set output format using ECMon, fixed
- c20m - sonic light flashing red/green, others green. not sending any data to dsm. → after setting output in ECmon it's giving data, but gas is all bad and sonic is mostly bad (occasional bursts of good data)
EC150: no gas from all the ones listed above, plus uw 10m. All the co2 values from the EC150s that are reporting are clipped in qctables.
...
Victrons: not staged, not planning to stage (actually want to stage at least one as part of a power drop test – Tony is building up power boxes for this)Ott: not yet staged, are we staging this?
Anything else I'm forgetting??