Installed S/N 1163, and 1164 at 30 and 16 meters, but witout data/power cables.
level | Licor 7500 S/N | level status status | |
---|---|---|---|
43m | 1167 43 m | port 17, working | |
30m | 1163 30 m | not cabled | |
16m | 1164 | 16 m | not cabled |
7m |
|
| |
2m |
|
|
DC power supply in power enclosure is rated at 3 A, which will not be large enough to drive system with 5 Licor 7500s, since each Licor needs about 0.9 A.
...