Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

According to the Licor 7500 manual, there is a 0.186 second sampling lag in the 7500. This can be increased with an output delay parameter, so that it can be synchronized with the sampling of a sonic. I've verified that the output delays on our 7500's are set to 0. Since the CSAT3 and Licor operate asynchronously, there is no advantage in further adjustment of the Licor delay to be a multiple of the CSAT3 sample period.

We have not been incorporating a 7500 lag in our data processing. We do correct for the 2 sample delay of the CSAT3 sonics. Subtracting 0.186 seconds from the 7500 samples does increase the correlation and fluxes between w and h2o and co2.

...