Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

According to the Licor 7500 manual, there is a 0.186 second sampling lag in the 7500. This can be increased with an output delay parameter, so that it can be synchronized with the sampling of a sonic. I've verified that the output delays on our 7500's are set to 0.

Since the CSAT3 and Licor operate asynchronously, there is I see no advantage in further adjustment of the Licor delay to be a multiple of the CSAT3 sample period.

...