Tom, Gordon arrived 4:19
Replaced dessicant in 4x radiometers.
Inspected mote radios with "dental" mirror. Radio in radiation mote (ID4) was tilted a bit out of its contacts on the board. Removed the mote and took it back to the base for maintenance. Did not have a spare mote that is programmed to talk to base mote #17.
Radiation mote battery has not been charging. Removed cover on rad mote battery box. One of the connectors was loose. Saw 16V on the PV side of the little circuit board.
Departed at 4:50