Feb 2, 2010: Patton, Golubieski, Maclean on site at 10:30 MST

Tasks:

  • Move Licor 7500 from 30 meters to 2 meters
  • Investigate why data system can't sample 4 7500's at 20 Hz
  • Measure guy tensions

First noticed a high-pitch sound from a TRH fan. Chris detemined it was the 16m TRH. The propeller had come off the shaft, and was not repairable. Swapped the 16m and 2m TRHs, so that now the 2m TRH does not have fan ventilation.

Initial licor 7500 serial port configuration:

port

height

status

2

2m

not connected

7

7m

OK, 9600 baud, 10Hz

17

16m

OK, 9600 baud, 10Hz

14

30m

OK, 9600 baud, 10Hz

11

43m

OK, 9600 baud, 10Hz

Looked at RS232 transmit signal from Licors with oscilloscope and a serial breakout box. Ports 17,11 and 7 all showed clean, square signals, ranging from -7V to +7V.

Moved 30m licor to 2m.

Looked at the signal ground of the 2m licor, with the oscilloscope probe ground connected to power ground. Saw hair-like spikes with a range of +- 1V at the bit frequency of the bits. By shorting signal ground to power ground we could get rid of these spikes.

Temporarily switched the cables on the 7m and 2m Licors, so that the 2m was plugged into port 7. Port 7 is on a Diamond Emerald serial card, which are the ports (5-20) that have not been working at 20Hz.

Up'd the sampling frequency of the 2m licor in port 7 to 20Hz, 19200 baud. Saw an increase in "spurious interrupts" which did not diminish when we shorted the signal and power grounds, so those spikes do not seem to be the issue.

Swapped dsm chassis boxes, replacing box #1 with box #8. Also swapped PC104 stacks. This configuration did not work well at all, with data_stats showing bad data rates. Swapped original PC104 stack back, and still saw problem. Eventually noticed that the input from the licor on port 17 had lots of jibberish. Moved this to port 20 and it cleaned up and the data_stats issues went away. So now believe the data_stats issue was due to a bad port 17 on the new box and not an issue with the pc104 stack.

The PC104 stack (which is what has been at the site all along):

card

SN

viper

#7, 3284

emerald 0, EMM-8M

W250017

emerald 1, EMM-8M

W237191

power

W248060

  • No labels