23-July
- changed online T=21.1 and i=162.6 mA to help locking to the 828.203 nm line at 16:00 MDT
20-July
- changed lab AC thermostat to 75 F -- the laser pointing fluctuations have been worse since it was running
18-July
- lab air conditioner was repaired during the day and left at 60F. At 20:45 MDT came in and moved it to 70F and peak up system
- iris was completely closed. Opened it ~1/10 of turn and near field range peaked up considerably
16-July
- changed online T=21.6 and i=177.4 mA to help locking to the 828.203 nm line at 15:30 MDT
- 200 counts out to 1800 meters range (system was untouched for 4 days)
11-July
- **one of the best sonde comparisons was at 11:00 PM last night -- good agreement from 1 km up to 5 km
- changed online T=21.7 and i=172.1mA to help locking to the 828.203 nm line
- removed code that obtained T, P, and RH from MSU from Labview
- installed latest Windows 7 updates and reseated 4GB memory
6-July:
- wavemeter was running intermittently -- having a hard time locking on the 828.203 nm line
- closed down the iris and peak up signal by adjusting the receiver fiber at 10:00 AM MDT
5-July:
- online changed to 828.203 nm at 10:00 AM MDT
- try to peak up signal with focus of telescope (little change)
- peak up signal slightly with adjustment to collimation (~1/2 turn)
2-July:
- AR coated (R~3%) window installed in the roof hatch. (Would add some angle and wavelength sensitivity -- not sure that it would be much)
- online was 828.201 but moved to 828.196 nm
- at 5:15 PM MDT adjusted the collimation to peak up the far field signal. Does not agree with collimation tester. Sagging mirror in gimbal mount?
- 200 counts out to 2500 meters range