QCR has leak ~3.5 mbar/min. Remove T from radome port and plug with thumb, passes leak test. Install 3/8 to 1/4" adapter onto radome port fails leak test. Snoop does not show leak. Remove tape and plug with rubber stopper, passes leak test.
Replaced T with PTFE. Nuts and ferrules left on hoses because ferrules have deformed hoses, will replace later.
CAVP_DPR & CAVP_DPL offset is different from previous cal. Since new slope is consistent on all data points, will consider cal good, just different.
RTH1 & RTH2, resistances don't auto-populate, missing bath cal file on ADS computer, referred to SEs
3 Comments
Unknown User (munnerly)
Pre cal test of cal cart. Capped off hoses, performed leak checks,
Test = +0.02 mbar/min
Ref = -0.002 mbar/min
Unknown User (munnerly)
QCR has leak ~3.5 mbar/min. Remove T from radome port and plug with thumb, passes leak test. Install 3/8 to 1/4" adapter onto radome port fails leak test. Snoop does not show leak. Remove tape and plug with rubber stopper, passes leak test.
Replaced T with PTFE. Nuts and ferrules left on hoses because ferrules have deformed hoses, will replace later.
CAVP_DPR & CAVP_DPL offset is different from previous cal. Since new slope is consistent on all data points, will consider cal good, just different.
RTH1 & RTH2, resistances don't auto-populate, missing bath cal file on ADS computer, referred to SEs
Unknown User (munnerly)
RTH1 & RTH2 auto-populate problem fixed by SEs. Calibrations complete with copies of screenshots sent to PMs, Adriana, Cowan and Kurt.