Ran an auto A2D calibration. All seems good. Channels 5 and 6 in Nose DSM analog card 1706 are a little high and this is a known issue.
CDP sent to DMT. Returned from DMT with a fixed loose pin and some new connectors. Put on plane and did not see CDP. Taken to the lab and was working. Loopback all 4 serial channels on LWO CPU. RS422 channel was bad. Swapped CPU and CDP works.
Spare server moved to R1.
VCSEL installed. Catherine adjusted the data system for a 19200 baud rate. It was sent to Southwest Sciences and they updated the firmware which created a different serial stream. Some data is there. No dewpoint. Goal is to straighten that out before the departure to Costa Rica.
1 Comment
Kyle Holden
Ran an auto A2D calibration. All seems good. Channels 5 and 6 in Nose DSM analog card 1706 are a little high and this is a known issue.
CDP sent to DMT. Returned from DMT with a fixed loose pin and some new connectors. Put on plane and did not see CDP. Taken to the lab and was working. Loopback all 4 serial channels on LWO CPU. RS422 channel was bad. Swapped CPU and CDP works.
Spare server moved to R1.
VCSEL installed. Catherine adjusted the data system for a 19200 baud rate. It was sent to Southwest Sciences and they updated the firmware which created a different serial stream. Some data is there. No dewpoint. Goal is to straighten that out before the departure to Costa Rica.