10/2/13, TWH:

Cleaned krypton and radiometers.

Vmote.rad reported low, but battery properly wired and measures 12.2 V both at the battery terminals and the battery connector on the mote.

Unfortunately, there were a couple of wisps of smoke when I tested the voltage in the mote and afterword the mote was not sampling properly (based on lights).

Left site at 13:25.