  1. Ran an auto A2D calibration. All seems good. Channels 5 and 6 in Nose DSM analog card 1706 are a little high and this is a known issue.


    CDP sent to DMT. Returned from DMT with a fixed loose pin and some new connectors.  Put on plane and did not see CDP.  Taken to the lab and was working.  Loopback all 4 serial channels on LWO CPU.  RS422 channel was bad.  Swapped CPU and CDP works.  


    Spare server moved to R1.


    VCSEL installed.  Catherine adjusted the data system for a 19200 baud rate.  It was sent to Southwest Sciences and they updated the firmware which created a different serial stream.  Some data is there.  No dewpoint.  Goal is to straighten that out before the departure to Costa Rica.