Staff: Jacquie, Josh

Weather: Sunny, light winds from the south, high 70's.

Notes: 

  • We replaced the bad CSAT's at t1 and t22 with the remaining spares.
  • We replaced the TRH casing at t41 (the ladder is now parked there)
  • The radiometers have been cleaned. 


t1CSAT3A SN 1312EC100 SN 1800PPE000020722



t22CSAT3 SN 0539PPE000016613



t41

  • Replaced the TRH casing. The SHT sensor is the original.


The Shelf of Bad Sensors

All the bad sensors have been labeled. 


One more thing: we noticed that the glass on the bad t1 CSAT3 was broken.





1 Comment

  1. Please, never set a sonic anemometer on the bench as in the first photo.  The sonic in the 6th photo is fine.  The data from a sonic anemometer critically depends on the geometry of the transducers (a 1mm change makes it useless) and we never want to apply force (including the weight of the sonic) on them.