- Most sonics are reporting NA, though they did work initially. Assuming this is ice related (try more heating?), but might check with Campbell to see if a configuration change would help.
- Most EC150s also show missing data. However, in at least one case, I've found the EC150 data to be okay – just missing because statsproc processes it in a group with the sonic data that ARE bad. Hopefully, fixing the sonics will bring back the EC150 values.
- A thermocouple appears to have been plugged in at sh on 19 Dec, but seems to have died on 21 Dec – just 2 days later. While it was plugged in, the data were okaylook good, though I haven't yet looked at spectranot exactly like tc (maybe a good thing). I do see noise above 3Hz during the day and 8Hz at night.
- Rfan.3m.prs is bad (T/RH ok). Housing should be replaced?
- Tsoil.mw is all -273. Hopefully, just a loose Binder connector.
- HRXL.cc stopped reporting 24 Dec.
...